START UP AND SHUT DOWN SEQUENCE
Start up:
1. Connect unit to a clean, filtered water source.
2. Connect air / compressor / nitrogen source.
3. Set the regulator to 80 PSI.
Note: When using a Nitrogen tank / cylinder, handpiece performance will diminish when the
tank pressure drops below 500 PSI. It will be necessary to switch out the tank / cylinder.
Shut down:
1. Clean all air handpieces according to the original manufacturer instructions. Thoroughly wipe
the unit, all handpieces and cables with a mild cleaning solution or disinfectant and a damp
cloth. Follow the procedures approved by your institution or use a validated infection control
procedure. Do not allow fluids to enter the chassis. Do not autoclave the main unit.
2. Remove and sterilize three way syringe tip (tip only), prophy angle, burs, low and high speed
handpieces. Prophy angle, low and high speed handpieces must be lubricated regularly
and after autoclave.
3. Disconnect the air / nitrogen source.
4. Disconnect unit from the water source.
.
CLEANING HANDPIECES, FOOTSWITCH AND POWER CABLES
After each procedure, or at least once a day, it is suggested that the handpieces and cables be
thoroughly cleaned and sterilized.
Clean the outer surface of the handpieces and cables with an antiseptic soap, rinse with water and
sterilize with a chemical sterilization solution.
Caution: No chemicals or cleaners should ever be used inside or allowed to get into the scaler
handpiece. Flush the handpiece thoroughly and completely with clean water.
The footswitch and power cables should be cleaned regularly by spraying a fine mist of sterilization or
cleaning solution on the cables. It should remain on the cables for the length of time recommended by
the manufacturer. Wipe the surface with a damp cloth and dry the cables completely.