OPTIONAL ACCESORIES
MAINTENANCE FREE PROPHY ANGLE
P-MF Maintenance free prophy angle
P-106 screw on rubber cups 144/pkt
CARE AND STERILIZATION PROCEDURES
After each prophy:
Rinse abrasive paste from head and cup area with water. Then remove cup.
Thoroughly clean the outside of angle with disinfectant.
Place angle into a sterilization bag
Follow sterilizer manufacturer’s recommendations.
Do not exceed 275 º F(135 ºC).
Keep angle in bag until ready for use.
You are now ready for your next prophy.
CAUTIONS AND WARNINGS:
Sterilize prior to disposal
Do not attempt to disassemble.
DO NOT submerge in liquids, including ultrasonic solutions.
Operate handpiece in the forward direction (counterclockwise when facing you) to prevent threaded
prophy cups from unscrewing during the procedure.
Not recommended for use above 3000 R.P.M.
If the head of the angle becomes hot during use, lubricate the rim of the cup / screw hole with mineral
oil.
Do not heat over 275 ºF (135 ºC).
Use only Engler Care Free Prophy Rubber Cups. Other brands will not properly seal the angle,
causing premature wear and voiding the warranty.
Use 1 year, sterilize, then dispose of properly.
YOUR CAREFREE ANGLE IS WARRANTED AGAINST DEFECTS IN MATERIAL AND
WORKMANSHIP FOR 6 MONTHS. A COPY OF OUR INVOICE OR PICKING TICKET WILL BE
REQUIRED AS PROOF OF PURCHASE.
CAREFREE ANGLE WARRANTY IS VOID IF
Engler Care-Free rubber cups are not used exclusively.
Sterilization procedure is not followed properly.
The angle has been submerged in any liquid.
The angle has been damaged or abused.
Damaged due to use at high speed.
26