23
If damaged, do not use this product!
Do not use in the presence of combustible or explosive materi-
als!
Never lay this product on the patient or in his/her direct vicinity.
Protect this product from any form of mechanical damage! Do
not throw! Do not use force! If there is a cable, and unless stated
otherwise, do not kink it or wrap it around the accompanying
product!
Set down the product carefully. Do not drop or throw.
Avoid tissue contact with parts that are not insulated in the joint
area of the instrument. Otherwise, tissue in the area of the joint
can suffer thermal damage during activation.
Erbe Elektromedizin expressly warns against modifying the
product. Any modification exempts Erbe Elektromedizin from
any and all liability.
6 How to use
6.1 Only use sterile products
Clean, disinfect and sterilize these products before using them
for the first time and before every subsequent use.
6.2 Checking the product
1. Check the product for damage, including the insulation or
coating.
If damaged, do not use this product!