Captair Flow Smart
Mobile ductless ltering clean air enclosure
16
Replacing the filters
Your device is equipped with FLEX™ lter technology that was congured to the sample’s protection needs when the device
was purchased. The conguration of the lter column is dependent on the applications carried out in the enclosure. These
applications may change over time. Your FLEX™ lter technology can therefore be recongured if your workstation is used
for applications other than those anticipated when the device is rst set up. If so, please contact us so that we can verify that
the current conguration is safe or it needs to be recongured.
.
The table below summarises all possible Flex™ lter technology congurations for your device
Captair Flow 321 - 391 -Smart models
Molecular lter Particulate lter Pre-ltre
Column
Conguration
1P - x1 x1
1C1P x1 x1 -
Captair Flow 483 - 714 -Smart models
Molecular lter Particulate lter
Pre-ltre
Column
Conguration
483 714 483 714 483 714
1P - - x3 x4 x3 x4
1C1P x3 x4 x3 x4 - -
UP
The table below summarises the different types of lters that Erlab® offers along with their elds of application.
Type AS
For organic vapors
HEPA H14
For particles
ULPA U16
For particles