Captair Flow Smart
Mobile ductless ltering clean air enclosure
22
Cleaning and maintenance
Mechanical item checks
Hinges: Hinges must be properly attached and must allow any items on the front side of the workstation to be instantly
and easily lifted upward.
Acrylic Parts: Ensure the transparency of the panels is a part of regular maintenance for the enclosure.
Cleaning the enclosure
The enclosure must be cleaned on a regular basis. This may be accomplished in several ways:
With soapy water followed by rinsing with clear water and drying with a soft, non-abrasive absorbent paper towel.
Or with a commercial pH neutral product followed by drying with a soft, non-abrasive, absorbent paper towel.
Or with a non-alcoholic disinfectant detergent.
Coated metal parts
These must be inspected and free from any traces of corrosion.
Check that there is no stagnant water in the spill tray.
Clean the spill tray if necessary.