21
When reassembling the ltration column, be sure that the fan module is positioned correctly to access your connection points.
ExamplE: Captair Flow 483 Smart
ExamplE: Captair Flow 321 Smart
A i r Fl o w
Reconnect the power supply to the fan
Recommendations for storing and using the filters
New lter shelf life and storage conditions:
New activated carbon molecular lters must be stored at in their original packaging at a temperature of between +10°C / 50°
F and +50°C / 120° F and a humidity level < 85% RH.
HEPA H14 and/or ULPA lters must be stored upright and kept dry.
If these conditions are respected, the maximum time a lter can be stored before use depends on the type of lter:
AS type carbon molecular lter 1 year from the date of delivery
HEPA/ULPA lter 2 years from the date of delivery
If the maximum shelf life is reached, we recommend not to install lters.
Filters replacement frequency
Prelters(s) should be replaced every year.
HEPA/ULPA lters should be replaced every 3 years on average.
Molecular lter should be replaced every 2 years on average.
Replacement frequency of carbon filter saturation
detection (Molecode option)
Molecular lters saturation detector – Molecode S (solvents) is installed as an option in your unit and use a semi-conductor
that must be changed every 5 years.
After this period, Erlab cannot guarantee performances stability nor detection sensitivity.
Equipment replacement requires the intervention of a qualied engineer, able to replace internal components and perform
system new conguration. The change requires less than one hour.
Erlab Maintenance department and accredited partners can provide this service. (service not available in all countries).
For further information, please get in touch with your contact person or Erlab dealer/distributor or visit our website.
Admissible weights on worksurfaces
The chart below reects loads that are evenly disributed across the work surface.
Trespa® Top LabPLUS
work top
Stainless steel
work top
Captair Flow 321 Smart 35 kg 35 kg
Captair Flow 391 Smart 50 kg 45 kg
Captair Flow 483 Smart 80 kg 80 kg
Captair Flow 714 Smart 115 kg 115 kg