B2 WEIGHING INDICATOR
6 Appendix
6.1 Error messages
Message Meaning Solution
#****$
Overload. Platform is loaded beyond
maximum capacity + 9e.
Unload platform.
%____&
Underload. Weight is less than zero setting
limit. Deadload may be changed.
Check the platform.
hata 1
Error in non-volatile parameter memory.
Weighing parameters/adjustment data are
missing or corrupted.
Configure the indicator. If the error
repeats call service.
hata 2
Possible hardware failure in analog circuit. Call service.
hata 3
Signal error. Possible disconnection or
erroneous connection of load cell cable.
Check load cell connection.
hata 4
Initial zero setting error. Weight value is
above or below zero setting range.
Check the platform.
hata 92
(when performing an adjustment)
Calculations show that load cell signal may
be out of range for weighings near capacity
(max).
The adjustment will be accepted, this
is just a warning.
hata 99
(when performing an adjustment) Calculated
load cell signal per scale interval is too low.
Check weighing parameters and/or
adjustment weight.
Press any key to close error message displays.
Note: Some errors can not be cancelled until the cause of error disappears. (e.g. hata 3)