B2 WEIGHING INDICATOR
1.1 Attention Before Start
Always use Earthed/Grounded Mains socket.
Do not connect powerful electrical machines on the same mains plug.
Disconnect mains connector before opening the enclosure.
Recommended load cell cable cross section is 1 mm². If the cable length is too long, use multiple
wires for ±Excitation connections.
If the sense inputs on the load cell connector are left empty, the device will display Err 3. Make sure
all the pins are connected.
Shielded twisted-pair cables must be used for load cell and communication.
For best accuracy, the device shall be powered on at least 30 minutes before adjustment.
Device setup shall be performed by authorized people.