EasyManuals Logo

Esco Class II User Manual

Esco Class II
73 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #25 background imageLoading...
Page #25 background image
15
Class II Biological Safety Cabinets
o 1N Hydrochloric Acid
o 1N Sodium Hydroxide
o 1% Quaternary Ammonium Compound
o 5% Formaldehyde
o 5,000 ppm Hypochlorite
o 2% Iodophor
o 5% Phenol
o 70% Ethyl Alcohol
Adequate contact time should be observed for effective decontamination and the time required
depends on the disinfectant agents, the concentration and the object of disinfection.
There is no one disinfectant agent that works with all organisms. Therefore, users and safety
professionals should carry out risk assessment to ensure that appropriate disinfectant agents and
validated decontamination procedures are used in decontaminating the BSC.
3.7. Gaseous Decontamination
Decontamination may frequently be carried out by means of formaldehyde fumigation or using other
decontamination agents, such as chlorine dioxide or hydrogen peroxide. Decontamination process should only
be carried out by qualified personnel.
In any of the following eventualities, the user should ensure that the BSC has been properly decontaminated,
keeping in mind the nature of the pathogens used:
At the time of moving/relocating the BSC
At the time of changing the type of work being carried out in the BSC
Before accessing contaminated areas for servicing (e.g. when filter needs replacement)
Periodically and as mandated by your risk assessment
3.8. Further Information
A Guide to Biosafety and Biological Safety Cabinets can be downloaded from
http://escoglobal.com/resource.php?id=13
An educational video on “Working Safely in your Biological Safety Cabinet” is available for viewing at
http://www.youtube.com/watch?v=ZnUW1N-JJz8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Esco Class II and is the answer not in the manual?

Esco Class II Specifications

General IconGeneral
LightingLED
UV LightOptional
CertificationNSF/ANSI 49, EN 12469
Airflow PatternLaminar
Inflow Velocity0.38 m/s ± 20%
Downflow Velocity0.25 m/s ± 20%
HEPA Filter Efficiency99.99% at 0.3 μm
MaterialStainless steel
Noise Level< 65 dBA
Front Opening200 mm (8")
ApplicationMicrobiology, cell culture
TypeBiosafety Cabinet