AFETY CONSIDERATION
Nevertheless, special precautions need to be observed:
The use of accessories and cable other than those specified, with the exception of accessories and cables
sold of the slit lamp microscope as replacement parts for internal components, may result in increased
EMISSIONS or decreased IMMUNITY or decreased LIFESPAN of the slit lamp microscope.
The slit lamp microscope should not be used adjacent to or stacked with other equipment. In case
adjacent or stacked use is necessary, The slit lamp microscope should be observed to verify normal
operation in the configuration in which it will be used.
The length of cables or cords must be greater than 1.8 meters.
a. Electromagnetic emissions
This product is intended for use in the electromagnetic environment specified below. It is up to the
customer or the user to verify that the instrument is used in this environment.
EMISSIONS TEST
COMPLIANCE
ELECTROMAGNETIC ENVIRONMENT GUIDELINES
Electromagnetic emissions (CISPR 11)
Group 1
The product uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
Disruptive voltage at power stations
(Conducted emissions) (CISPR 11)
Class A
The product is suitable for use in all
establishments other than domestic and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes.
Harmonic current emission (IEC61000-3-2)
Class A
Voltage variations, voltage fluctuations and
flicker (IEC61000-3-3)
Complies
b. Magnetic and electromagnetic immunity
The product is intended for use in the electromagnetic environment specified below. It is up to the
customer or the user to verify that the instrument is used in this environment.
IMMUNITY TEST
TEST LEVEL IEC
60601
COMPLIANCE LEVEL
ELECTROMAGNETIC ENVIRONMENT
GUIDELINES
Electrostatic discharge (ESD)
(IEC61000-4-2)
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the
relative humidity should be at
least 30%.
Electrical fast transients and
bursts (IEC61000-4-4)
± 2 kV for power
supply lines
± 1 kV for the
signal ports
± 2 kV for power
supply lines
± 1 kV for the signal
ports
Mains power quality should be that
of a typical commercial or hospital
environment
Surge (IEC61000-4-5)
± 1 kV lines to
lines
± 2 kV lines to
earth
± 1 kV lines to lines
± 2 kV lines to earth
Mains power quality should be that
of a typical commercial or hospital
environment.
Power frequency (50/60Hz)
magnetic field
(IEC61000-4-8)
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment
< 5% U
T
< 5% U
T
Voltage dips, short interruptions
and voltage variations (IEC61000-
4-11)
(>95% dip in U
T
)
For 0.5 cycles
40% in U
T
(60% dip in U
T
)
For 5 cycles
(>95% dip in U
T
)
For 0.5 cycles
40% in U
T
(60% dip in U
T
)
For 5 cycles
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
product requires continued
operation during power mains
interruptions, it is recommended
SL650L - Slit lamp microscope > V1 - 10-2021 - CE
33

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Essilor Instruments SL 650 and is the answer not in the manual?

Essilor Instruments SL 650 Specifications

General IconGeneral
BrandEssilor Instruments
ModelSL 650
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals