U
SER MANUAL
> S
PECIFICATIONS
1. Technical data
The expected life of the device and its components is 8 years.
Microscope
Type: Galilean type
Magnification change: Five Magnifications
Eyepieces: 12.5X
Angle between eyepieces: 10°
Total magnification ratio: 6.3X, 10X, 16X, 25X, 40X
Pupillary adjustment: 52 mm ~ 80 mm
Diopter adjustment: -8D +8D
Field of view:40X (5.7 mm), 25X (8.9 mm), 16X (14 mm), 10X (22.3 mm), 6.3X (36.2 mm)
Slit width: Continuously variable from 0 to 14 mm (at 14 mm, slit becomes a circle)
Slit length: Continuously variable from 1 mm to 14 mm
Aperture diameters: 14 mm, 10 mm, 5 mm, 3 mm, 2 mm, 1 mm, 0.2
mmSlit angle: - 180° (Adjustable both vertical and horizontal)
Slit inclination: 4 steps: 5°, 10°, 15°, 20°
Filters: Heat-absorbing filter, ND filter, Red-free, Cobalt blue filter, Build-in Yellow Filter
Lamp: 3V / 3W LED module
Luminance: 150 Klx
Base
Longitudinal movement: 115 mm
Lateral movement: 110 mm
Vertical movement: 30 mm
Fine base movement: 15mm
Chinrest
Vertical movement: 80 mm
Power
Input: 100~240V, 50/60Hz
Rated current: 1.2A
Output :
Lamp: 3V
Fixation: 15V 20mA
Dimension and weight
Dimension: 770 mm x 470 mm x 570 mm
Gross weight: 26kg
Net weight: 18 kg
SL650L - Slit lamp microscope > V1 - 10-2021 - CE