4.Calibrate Exp. Valve
Turn the Knob to select Exp Valve
Cal and press it, see right figure.
Calibrating procedure refer to
Calibrating pressure sensor.
The other operation like the press
sensor calibrate.
NOTE: Before calibrating
expiratory valve, ensure no
leakage!