Press .
Press
or to select the desired value (0.00 to 10.0).
Press
to confirm.
6.9 P3.6 t.nr (Normalization Temperature in ºC)
When Automatic Temperature Compensation is used, measurements
are ad
justed by the temperature coefficient to the normalization
temperature. The default value is 25 ºC.
Press
.
Press
or to select the desired value (15.0 to 30.0).
Press
to confirm.
6.10 P3.7 ACAL (Automatic Conductivity Calibration)
See Section 4.1 for more details on Automatic Conductivity Calibration.
.
Press
Press
or to select “Yes” (Automatic) or “No” (Manual).
Press
to confirm.
6.11 P3.8 SPC (Single Point Calibration)
See Section 4.2 for details on Single Poi
nt & Multi Point Calibration.
.
Press
Press
or to select “Yes” (Single) or “No” (Multi).
Press
to confirm.
6.12 P3.9 CELL (Nominal Cell C
onstant)
The CON 700 includes a probe with a nominal cell constant (k) of 1.0.
Use probes with k = 0.1 and 10 (sold separately) for improved
11