11. SPECIFICATIONS
Conductivity
Range
..µS to 200.0 mS
Conductivitity
Resolution
0.01 / 0.1 / 1 μS ;
0.01 / 0.1 mS
TDS Range .. to 100.0 ppt @ 0.5 fact (200.0 @ 1 factor)
Resolution 0.01 / 0.1 / 1 ppm ; 0.01 / 0.1 ppt
Accuracy ±1% Full Scale
Cell Constant 0.1, 1.0, 10.0 (selectable)
TDS Factor 0.40 to 1.00 (adjustable)
Calibration points
Automatic (4 points); Maximum 1 per range
Manual (5 points
); Maximum 1 per range
Temperature
Range (Meter)
0.0 to 100.0 ºC / 32.0 to 212.0 ºF
Compensation
Type & Range
Automatic with supplied cell or Manual.
0.0 to 100 ºC / 32.0 to 212.0 ºF,
0.0 to 80 ºC / 32.0 to 176.0 ºF with supplied cell
Resolution
0.1 ºC / 0.1 ºF
Accuracy
± 0.5ºC / ± 0.9ºF
Calibration
Offset in 0.1 º increments; Offset range: ± 5.0 ºC / ± 9.0 ºF
Coefficient
0.0 to 10.00 per ºC
Auto-ranging Yes
Hold Function Yes
Memory 100 data sets
Input 8-pin DIN
Power AC/DC 9V, 6W Adapter (100/240 VAC, 50-60Hz)
Dimensions (mm) meter only = 175 (L) x 155 (W) x 69 (H)