performance in extreme sample ranges. Use this setup function to
change the cell constant if necessary. Meter default is 1.0 to match the
included probe.
k = 0.1 ideal for low measurements <20 µS (<10 ppm).
k = 1.0 ideal for mid-range measurements
k = 10 ideal for high measurements >20 mS (>10 ppt).
Press
.
Press
or to select 0.1, 1.0, or 10.0
Press
to confirm.
6.13 P4.0 rSt (Reset)
Press
.
Press
or to select “Yes” (Reset) or “No” (Cancel).
If “Yes”, press
or to select “Cal” (calibration reset
only) or “FCt” (complete reset to factory default settings).
Press
to confirm.
6.14 P5.0 CLr (Clear Memory)
Press
.
Press
or to select “Yes” (Erase memory) or “No”.
Press
to confirm.
12