CONDUCTIVITY PROCEDURES
RINSE METER SENSORS
Rinse the sensors in clean water and blot dry
(never
rub or touch the pH glass bulb)
.
1
TEST PROCEDURE
VERIFY CALIBRATION
After calibration, dip the meter in 12.88 mS/cm
calibration solution. If the value is accurate, it is not
necessary to conduct a 2nd point calibration. If it is
inaccurate, follow steps 5 to 6 to complete the 2
nd
point
calibration using the 12.88 mS/cm buffer solution.
8
RUN CALIBRATION
Dip the meter in 1413 μS/cm calibration solution, stir
gently, and allow the probe to sit in the solution until
a stable reading is reached. When the
stable
icon appears on the LCD, press the key to
complete one-point calibration. The meter returns to
measurement mode and the icon will appear at
the bottom left of the LCD.
6
TEST
Fill Sample Collection Bottle with sample to be tested.
Submerge the probe in the sample and stir gently.
Then allow to sit in the bottle.
3
SELECT MENU
Press and re-press the button to select until
Cond appears in the top left corner of the display.
2
RINSE METER SENSORS
Rinse the sensors in clean water and blot dry
(never
rub or touch the pH glass bulb)
.
7
READ RESULTS
Record the readings after the stable icon comes up.
Press to switch from Conductivity to TDS and
Salinity. When fi nished with readings, rinse the probe
with clean water and return it to the included storage
solution or to pH 4 buffer. NOTE: If using the eXact
iDip
®
app, press the Save Result button at the top of the
Bluetooth Test page to store the displayed result.