• The TDS and Salinity measurements are calculated from the
conductivity measurements by using conversion factors.
The meter can recognize 84μS, 1413 μS/cm and 12.88 mS/cm
conductivity calibration solutions. The user can conduct 1 to 3 points
calibration. Refer to the table below. Usually calibrating the tester with
1413 μS/cm conductivity buffer solution alone shall meet the testing
requirement. However, we recommend choosing a calibration solution
close to the expected conductivity of your sample.
Calibration indication
Icon
Calibration
Standards
Measuring Range
84μS/cm 0 - 200 μS/cm
1413 μS/cm 200 - 1999 μS/cm
12.88 mS/cm 2 - 20 mS/cm
The tester has been calibrated before leaving the factory. For best
accuracy we recommend testing a conductivity buffer solution
before use. If the error is large, then calibration is needed.
• To maintain the accuracy of conductivity calibration solutions, we
recommend replacing the solutions after 3 to 5 uses. To prevent
contamination, Do NOT pour used calibration solutions back into
the original bottles.
• Temperature compensation factor: The default setting of the
temperature compensation factor is 2.0%/°C. Users can adjust the
factor depending on the solution or experimental data (see table
below). Use parameter setting P4 to adjust the factor.
Solution
Temperature
compensation factor
NaCl
2.12%/°C
5% NaOH
1.72%/°C
Dilute ammonia
1.88%/°C
10% Hydrochloric acid
1.32%/°C
5% Sulfuric acid
0.96%/°C
• 1000 ppm = 1 ppt 1000 μS = 1 mS
• TDS and conductivity are linearly related; the conversion factor
ranges from 0.40-1.00. The factory default setting is 0.71. Salinity
and conductivity are also linearly related; the conversion factor
is 0.5. The tester only needs to be calibrated in Conductivity
mode. After calibration for conductivity, the meter can switch from
conductivity to TDS or salinity.
Conversion Example: if conductivity measurement is 1000μS/
cm, then the default TDS measurement will be 710 ppm (under the
default 0.71 conversion factor), and the salinity will be 0.5 ppt.
• Avoid cross-contaminating the conductivity standard solutions.
Rinse the sensors and blot dry after every measurement.
CONDUCTIVITY TEST NOTES

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the eXact pH+ and is the answer not in the manual?

eXact pH+ Specifications

General IconGeneral
pH Range0.00 to 14.00 pH
Resolution0.01 pH
Waterproof RatingIP67
Temperature Resolution0.1 °C / 0.1 °F
Temperature Accuracy±0.5 °C / ±1 °F
CalibrationAutomatic, up to 3 points
Power Source4 x AAA batteries
Temperature CompensationAutomatic