CurrentMeasurementConsiderations:
Currentmeasurementmustbeperformedinthetemperaturerangeof0~40
o
C(32to104
o
F).
Whenpressingthetrigger,donotsuddenlyreleaseit;theclampissensitivetomagnetism,
heat,andmechanicalstressandsuchimpactwillcausethereadingtofluctuatebriefly.
IfthereadingispositiveinthemeasurementofDCcurrent,thedirectionofcurrentisfrom
toptobottom(meterfaceplateisthetopandthemeterbackisthebottom).
Inordertoensuremeasurementaccuracy,theconductormeasuredmustbepositionedin
themiddleoftheclampingarea;otherwiseadditionalerrorofreadingof±1.0%(typical)will
begenerated.
Theaccuracyofnon‐sinusoidal
wavesmustbeadjustedasfollows:
Crestfactor1~2,accuracyincreasesby3%
Crestfactor2~2.5,accuracyincreasesby5%
Crestfactor2.5~3,accuracyincreasesby7%