EasyManua.ls Logo

Extrel MAX300-RTG - Calibration Gases

Default Icon
87 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MAX300-RTG: Installation Manual Hardware Installation
37
COMPRESSED AIR INLET
FOR VALVE ACTUATION ONLY
PRESSURE: 40-100 PSIG (276-690 kPa)
FLOW: 0.1 CFH (2.8 L/H)
CLEAN DRY INSTRUMENT AIR
SAMPLE FLOW REQUIREMENTS
BALANCE FLOWS WITHIN A FACTOR OF 5
VALVE BODY PRESSURE MUST NOT
EXCEED 5 PSIG (34 kPa)
Figure 17: FastValve (40/80/120/160 port valve option) for use with
non-explosive sample
Calibration Gases
Calibration Gas Connections
The customer must connect the calibration gas cylinders to the MAX300-RTG
without contamination. Details of the connection may vary depending on the valve
option installed. Figure 18 below depicts the recommended hardware for
connecting calibration gas cylinders to the MAX300-RTG.
A common configuration is Valco rotary valves using 1/16” fittings. Each cylinder
of calibration gas requires:
High quality, low outlet pressure, two stage regulator with a stainless steel
diaphragm and appropriate CGA fitting. Please contact your gas supplier
for special considerations such as corrosive gases.
1/16" O.D. x 0.030" I.D. stainless steel tubing. Length as required to route
plumbing from MAX300-RTG to the gas storage area.
1/16" Valco nuts, ferrules, and plugs (supplied with the valve kit).
COMPRESSED AIR INLET
FOR VALVE ACTUATION ONLY
PRESSURE: 60 100 PSIG (414 690 kPa)
FLOW: 0.1 CFH (2.8L/H)
CLEAN DRY INSTRUMENT AIR
SAMPLE FLOW REQUIREMENTS
BALANCE FLOWS WITHIN A FACTOR OF 5
VALVE BODY PRESSURE MUST NOT
EXCEED 5 PSIG (34 kPa)