EasyManua.ls Logo

FBD 372 - Initial Power up

FBD 372
64 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
37X INSTALLATION & OPERATION MANUAL
372/373 MODELS
13
Water Flow
Control
Syrup Flow
Control
5.3 Water Flow Rate and Brix
The water flow rate and brix for each barrel must be verified and adjusted for optimal performance. Use the following
procedure to adjust the water flow rate and set the brix for each barrel:
For multi-flavor dispensers, the flavor injection modules will also need to be calibrated. Please refer to the Multi-Flavor
Merchandiser Manual (24-M4MF-0001) for the calibration procedure.
The target brix for most sugar flavors is between 13.5 and 15. The target brix for most “light” flavors will be lower, usually
between 9 and 10. Check the specifications for the flavors if unsure of the target brix
Note: Light or diet products have a low brix and can cause problems with the dispenser if not properly set up. Although
these products contain ingredients that ensure the sugars freeze, they can still create excessive ice if not carefully brixed
according to their specifications. Low brix issues will be associated with events such as audible ice scraping during freeze
and low beater error problems.
1 Turn the syrup shuto valve to the OFF position (Figure 5.2).
2
Adjust the water ow rate to 1.5 oz/sec (44.4 ml/sec) by dispensing water from the solution module
through the sample port for 10 seconds into a measuring cup. The amount measured should
be 15 oz (444 ml). Turn the ow control clockwise to increase the ow and counterclockwise to
reduce the ow.
Note: The WATER AND BRIX CALIBRATION screen under the MAINTENANCE MENU can be used to
help with timing.
3
Turn the syrup shuto valve to the ON position and dispense a cup of uid from the sample port
to establish a stable ow.
4
Once a good water and syrup mixture is obtained, dispense a 16 oz sample and measure the
sample brix with a refractometer. Be sure to thoroughly mix the sample rst by pouring back and
fourth between two cups.
5
Turn the syrup ow control counterclockwise to decrease the brix. Turn the syrup ow control
clockwise to increase the brix.
6 Dispense and discard several samples after adjusting the brix.
Sample Port
Sample Valve
(closed position
shown)
Water Shutoff
Valve (open
position shown)
Syrup Shutoff
Valve (open
position shown)
Turning Clockwise
Increases Flow Rate
+
Turning Counterclockwise
Decreases Flow Rate
Flow Rate Adjustment
Figure 5.2 Solution Module

Table of Contents

Related product manuals