EasyManua.ls Logo

FCI ST50 - Pulse Output Wiring; Source Output Wiring; Sink Output Wiring

FCI ST50
17 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
FLUID COMPONENTS INTERNATIONAL LLC ST50/ST75 PDA SOFTWARE
Configuring and Setting up the Pulse Output Channels
The Pulse Output Channels are used to Provide Totalizer and Process Alarm information. Totalizer information is provided
as a voltage pulse, and Alarm information is provided as a Voltage “State Transition”; Low-To –High or High-to-Low.
There are 2 channels; Channel 1 is a Source Channel, and Channel 2 is a Sink Channel. (See ST50/ST75 Installation
Instructions for information on how to connect to these 2 channels).
From the Setup Menu, tap on the “Output Config” button to open the setup screen for the ST50/ST75 Analog Channel
selection, and the Pulse Output channel setup.
Tap on the “Pulse Out Setup” button to open the Pulse Output
Channels Configuration screen.
Tap On the “Get All” button to read the current setup of the
ST50/ST75.
Chose the signal to be assigned to the Source Channel.
Choose the signal to be assigned to the Sink Channel.
Select and enter the new values for “Pulse Factor”, and “Sample
Period”. In most cases these values do not do change from the
factory settings.
If the “Flow Alarm” has been selected, then also select the pulse
transition format Low-to-High or High-to-Low.
Select the Flow Alarm set point (It is selectable in SFPS).
Some Definitions
Sample Period: This is the “Pulse Output” channel update rate. Pulse output is not a continuous outputting signal, but a new
value is updated based on the Sample Period parameter.
Pulse Factor: This is the equivalency number to the corresponding instrument volumetric flow units. (i.e. 1 pulse is equal
to 1 NCMH for a factor of 1, if NCMH are the volumetric units selected.
This page is subject to proprietary rights statement on last page. 8 06EN003372 Rev. -

Table of Contents

Other manuals for FCI ST50

Related product manuals