Fluid Components International LLC ST75/ST75V Mass Flow Meter
This page is subject to proprietary rights statement on last page
24
06EN003368 Rev. G
Table 6. ST75/ST75V List of CLI Commands, Cont.
Command Line Password: 357
COMMAND
MNEMONIC
COMMAND
FUNCTION
COMMAND DESCRIPTION DATA TYPE
K[1-4] R/W K Factors Float
L0 R/W Line Size 0 Float
L1 R/W Line Size 1 Float
MN R/W Minow Float
MX R/W Maxow Float
NN R/W Namur Mode Integer
OM R/W Outmode Integer
PF R/W Pulse Factor Float
PL R/W Pulse Out Integer
PS R/W Pulse Sample Period Float
PW R/W Pulse Width Float
P0 R/W Switch Point0 Integer
P1 R/W Switch Point1 Integer
RO R/W RollOver Cntr Long
RR R Reference R Float
RS R/W RefR Slope Float
RF R/W RefR Offset Float
SF R SFPS Flow Float
SN R/W Serial Number String (16 chars max.)
SO R/W Shop Order Number String (16 chars max.)
S0 R/W SpanDAC0 for 4-20mA #1 Integer
S3 R/W SpanDAC1 for 4-20mA #2 Integer
S2 W Save FACTORY N/A
TC R TCDeltaR Float
TD R/W Tcslp Float
TF R/W Totalizer OFF/ON Flag Integer
TM R/W Cmintemp Float
TP R/W Totalizer Temperature Flag Integer
TT R/W Totalizer Value Float
TX R/W Cmaxtemp Float
TZ R Temperature Float
T0 R/W Tcslp0 Float
T2 R/W Tcslp2 Float
T3 R/W TSpanDAC0 for 4-20mA #1 Integer
T7 R/W TSpanDAC1 for 4-20mA #2 Integer
T5 R/W TZeroDAC0 for 4-20mA #1 Integer
T8 R/W TZeroDAC1 for 4-20mA #2 Integer
UF R User Flow Float
UK R User FlowK Float
UN R/W User Name String (16 chars max.)
VN R Version Number String (16 chars max.)
XX R/W Test Flow Rate (SFPS) Float
XY W Delete Test Flow Rat e Float
Z0 R/W ZeroDAC0 for 4-20mA #1 Integer
Z2 R/W ZeroDAC1 for 4-20mA #2 Integer