47
or
OPEN DOOR ALARM
COMPARTMENT A
or
OPEN DOOR ALARM
COMPARTMENT B
Check delay time for the
alarms. If necessary, change
them, Par 5.2.4
The door has been left open for
longer than the alarm delay set
(45 -60 seconds)
The door is not properly closed
If the problem occurs again,
contact technical service
or
PRODUCT LOW
TEMPERATURE
Check the set limits with
reference to the working
setpoint. See Par. 5.2.4
Too narrow high/low alarm limits
The refrigerator operates and
is controlled by the on / off of
the last cycles of the
compressor, but you need to
contact Technical Service
Tap on icon in order to
view alarm list and to mute it
(par.5.2.16)
Turn the refrigerator off from
the display, -STAND BY-
Par.5.2.8. Turn it on after a
few seconds.
If the problem occurs again,
do check the voltage by a
qualified electrician.
Contact sales department for
a voltage stabilizer.
220V or 115V mains is not stable
and registers spikes lower than
the minimum accepted value.
LOW EVAPORATION
TEMPERATURE
Turn the refrigerator off from
the display, -STAND BY-
Par.5.2.8. After a while turn it
on and the visual and
acoustic alarm is off.
Tap the off alarm/fault icon so
to permanently delete the
signal (par.5.2.16).
If the problem occurs again,
contact technical service
Evaporator fan is stuck
Electronic board relay fault
Ice formed on the evaporator
Faulty defrost heater
Unbalanced evaporator fan
Electrical or electronic faults