EasyManua.ls Logo

FIOCCHETTI ECT-F TOUCH - Page 51

FIOCCHETTI ECT-F TOUCH
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
51
PROBE FAILURE
S2 - S3
In case of S2 or S3 robe
failure, device operation does
not take place regularly.
Turn the refrigerator off from
the display, -STAND BY-
Par.5.2.8. After a while turn it
on and the visual and
acoustic alarm is off.
Tap the off alarm/fault icon so
to permanently delete the
signal (par.5.2.16).
Always contact technical
service for the probe
replacement, especially for
S2 probe.
Disconnected or faulty probe
S2 Evaporator probe failure
S3 Condenser probe failure
--%
Contact technical service for
probe replacement
Humidity viewer probe
disconnected or faulty
PROBE S1m DMLP
FAILURE
PROBE S2m DMLP
FAILURE
Device still works properly,
but it does no DMLP S1m
probe recording.
Turn the refrigerator off from
the diplay, -STAND BY-
Par.5.2.8. After a while turn it
on and the visual and
acoustic alarm is off.
Tap the off alarm/fault icon so
to permanently delete the
signal (par.5.2.16).
Contact service for probe
replacement
Disconnected or faulty probe
S1m DMLP probe failure
S2m DMLP probe failure (B
compartment or product
simulator)
UNBALANCED
PROBES ALARM
(with DMLP digital
monitor)
Turn the refrigerator off from
the display, -STAND BY-
Par.5.2.8. After a while turn it
on and the visual and
acoustic alarm is off.
Tap the off alarm/fault icon so
to permanently delete the
signal (par.5.2.16).
If the problem occurs again,
contact technical service
Different readings between S1
probe and S1m probe (DMLP
digital monitor)

Table of Contents

Related product manuals