14
Clave 23+ Instruction Manual v1.4
7. Maintenance
Frequency Maintenance operation
Daily
Clean the door seal
Clean the external surface
Drain the used water tank
Weekly
Clean the distilled water tank (drain tank)
Clean the sterilization chamber
Monthly Clean the filter in the chamber and tank
Every 3 months Replace the bacteria filter
Every 6 ~ 12 months Replace the door seal
7.1 Clean the distilled water tank every week with isopropyl alcohol or a medical
disinfectant. Drain unused water.
7.2 Clean the chamber weekly.
7.2.1 Remove all trays and the tray rack from the chamber.
7.2.2 Clean the chamber with a smooth cloth saturated with distilled water.
7.2.3 Apply the same procedure to the trays and rack.
7.3 Clean the door seal.
7.3.1 Clean the door seal weekly with a smooth cloth saturated with distilled water.