19
Clave 23+ Instruction Manual v1.4
Appendix 2
Diagrams of the sterilization programs
Programs
Temp
°C
Pressure
kPa
Holding
time
Total
time
(min)
Type
Max.
load
(kg)
load
per tray
Unwrapped
134 210 4 25~45
Unwrapped solid
material
5.00 1.50
121 110 20 30~60
Liquid 134 210 10 35~60 Liquid
1.20 0.40
Plastic 121 110 30 35~65 Plastic
Wrapped
134 210 6 35~65
5.00 1.50
121 110 20 35~55
Single wrapped solid
or hollow material
4.00 1.20
Textile
134
121
210
110
8
30
45~65
50~75
Unwrapped porous
material
1.25 0.40
Single wrapped
porous material
1.10 0.30
Dual wrapped porous
material
0.75 0.25
Single wrapped
hollow material
4.00 1.25
Dual wrapped solid
and hollow material
2.00 0.60
Prion
134 210 18 50~75
Unwrapped porous
material
1.25 0.40
Single wrapped
porous material
1.10 0.30
Dual wrapped porous
material
0.75 0.25
Single wrapped
hollow material
4.00 1.25
Dual wrapped solid
and hollow material
2.00 0.60
-- -- -- 1~20 -- -- --
The time required for the sterilizer to be ready for routine use after the power is switched on is
less than 5 minutes.
The maximum temperature of the 134 °C sterilization cycle is 137 °C.
The maximum temperature of the 121 °C sterilization cycle is 124 °C.