Access Array System for the Ion Torrent PGM Sequencing System: User Guide
Chapter 1: Access Array System Overview
Before You Begin
15
Before You Begin
IMPORTANT Before using the Access Array system, read and understand the detailed
instructions and safety guidelines in Appendix D.
Use good laboratory practices to minimize contamination of samples:
Use a new pipette tip for every new sample.
Whenever possible, separate pre- and post-PCR activities. Dedicate laboratory
materials to designated areas.
Unless otherwise specified, thaw reagents at room temperature (15–30 °C), and then use
them at room temperature.
Retrieve only the reagents required from each kit based on the number of IFCs you will
run.
Only use the reagents provided in the required kit.
Do not swap reagents between kit lots.
Mix and centrifuge reagents as directed.
Avoid creating bubbles when transferring reagents to the IFC.
This protocols in this document require the use of updated scripts for FC1 (v1.6 or later) and IFC
Controller AX (v2.8 or later) that can be downloaded from fluidigm.com/software.