4.4.3. TERMINAL CONNECTORS
For Intrinsically Safe applications: read chapter 5.
The following terminal connectors are available:
Fig. 9: Overview of terminal connectors standard configuration F110-P and options.
REMARKS: TERMINAL CONNECTORS:
Terminal GND- 01- 02: Power Supply - only available with type PD / PF or PM:
TYPE SENSOR SUPPLY
Terminal
backlight
TYPE AA
T
YPE
AU
T
ype
O
A
T
ype
O
R
GND 01 02
PD 8-24V AC 8,2 / 12 / 24V max. 50mA
AC AC
PD 8-30V DC 8,2 / 12 / 24V max. 50mA L- L+
PF
24V AC ± 15%
8,2 / 12 / 24V max. 400mA
AC AC
PF
24V DC ± 15%
8,2 / 12 / 24V max. 400mA L- L+
PM
115-230V AC ± 15%
8,2 / 12 / 24V max. 400mA EARTH AC AC
Note PD
do not use a AC autotransformer (Spartrafo) without a galvanic isolation.
Note PF / PM
The total consumption of the sensors and outputs may not exceed 400mA@24V
=option
Note: for power supply type PX: please read Terminal 07-08 !
78
I
GND 1
POWER SUPPLY
TYPE PD / PF / PM
NL1
2
I
910
SIGNAL
11
+
1265
R1 R1
13
RESET
SENSOR SIGNAL
TYPE P:
PULSE INPUT
EXTERNAL
RESET
TYPE IB
(OPTION)
PULSE
OUTPUT
TYPE
OA/ OT / OR
A
NALO
G
OUTPUT TYPE
AA / AB
AI / AP / AU
POWER SUPPLY
TYPE PX