10 Users Manual
16
Maximum Voltage Between any
Terminal and Earth Ground
600 V rms
Display 3 3/4-digits, 4000 counts, updates 4/sec
Operating Temperature 10 °C to 50 °C
Storage Temperature 30 °C to 60 °C indefinitely (to 40 °C for 100 hrs)
Temperature Coefficient 0.1 x (specified accuracy) / °C (<18 °C or >28 °C)
Relative Humidity
0 % to 90 % (10 °C to 35 °C);
0 % to 70 % (35 °C to 50 °C)
Battery Type 9V, NEDA 1604 or IEC 6F22
Battery Life Alkaline, 650 hrs continuous;
Carbon-zinc, 450 hrs continuous
Shock, Vibration
1 meter shock. Vibration per MIL-T-28800D for a Class 3
Instrument
Size (H x W x L) 1.35 x 2.75 x 5.55 inches; (3.46 x 7.05 x 14.23 cm)
Weight 10 oz (286 g)
Safety Complies with Protection Class II requirement of UL1244,
ANSI/ISA-S82, CSA C22.2 No 231, and VDE 0411, and
IEC 1010 overvoltage CAT III. CAT III instruments are
designed to protect against transients in fixed-equipment
installations at the distribution level.
EMI Regulations Complies with FCC Part 15, Class B, and VDE 0871B.