10 Users Manual
18
Function
Overload
Protection *
Input
Impedance
(Nominal)
Common Mode
Rejection Ration
(1 kQ Unbalance)
Normal Mode
Rejection
VF 600 V dc >10 MQ <100 pF
50 Hz or 60 Hz
100 dB @ dc @
50 or 60 Hz
>50 dB @
50 or 60 Hz
VB 600 V rms
>5 MQ <100 pF
(ac-coupled)
>60 dB, dc, 60 Hz NA
Full Scale Voltage Short Circuit
Current
Open Circuit
Test Voltage
To 4.0 M 40 M
600 V rms < 1.5 V dc <450 mV dc <1.5 V dc <500 µA
G 600 V rms 2.4 to 3.0 V dc 2.400 V dc 0.95 mA typical
* 3 x 10
6
V Hz Maximum