1550C/1555
Calibration Manual
2
How to Contact Fluke
To contact Fluke, call one of the following telephone numbers:
• Technical Support USA: 1-800-44-FLUKE (1-800-443-5853)
• Calibration/Repair USA: 1-888-99-FLUKE (1-888-993-5853)
• Canada: 1-800-36-FLUKE (1-800-363-5853)
• Europe: +31 402-675-200
• Japan: +81-3-3434-0181
• Singapore: +65-738-5655
• Anywhere in the world: +1-425-446-5500
Or, visit Fluke's website at www.fluke.com
.
To register your product, visit http://register.fluke.com
.
To see, print, or download the latest manual supplement, visit
http://us.fluke.com/usen/support/manuals
.
Safety Information
A Warning identifies conditions and actions that pose hazard(s) to the user; A Caution
identifies conditions and procedures that could cause Product damage, equipment under
test damage, or permanent loss of data.
Warning: Read before product use.
To prevent possible electrical shock, fire, or personal injury:
• Read all safety Information before you use the product.
• Before and after testing, confirm that the product does not
indicate the presence of a hazardous voltage, see Figure 2.
If the product beeps continuously and a hazardous voltage
is shown on the display, remove power from the circuit
under test or allow the installation capacitance to fully
discharge.
• Use the product only as specified, or the protection supplied
by the product can be compromised.
• Connect the common test lead before the live test lead and
remove the live test lead before the common test lead.
• Do not disconnect the test leads before a test has been
completed and the test voltage at the terminals has returned
to zero. This ensures that any charged capacitance is fully
discharged.
• Disconnect power and discharge all high-voltage capacitors
before you measure resistance or capacitance.
• Do not work alone.
• Do not use the product around vapor, or in damp or wet
environments.