Fluke 1621 - Maximum Deviations; RE Resistance Measurement

Fluke 1621
26 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1621
Users Manual
16
Temperature
Coefficient:
± 0.1 % of range per degree Kelvin
Safety: IEC/EN 61010-1, 600VCATII, pollution degree 2
Maximum Deviations:
Parameter Influence Factor
Deviation
influence
E
1
Position 0 %
E
2
Supply Voltage 0 %
E
3
Temperature E
3
2.3 %
E
4
Serial Interference Voltage (20 V) 0.6 %
E
5
Probe- and Auxiliary probe
Resistance
10 %
Test voltage: 3.7 kV
Protection type: IP 40; IEC/EN 60529
Electromagnetic
compatibility:
Emission: IEC/EN 61326 Class B
Immunity: IEC/EN 61326 Annex C
Dimensions: 113 x 54 x 216 mm (4.5 x 2.1 x 8.5 in), including
holster
Weight: 850 g (1.9 lb), including standard accessories,
Volume approximately 600 cm
3
R
E Resistance Measurement
Measuring method: Current-voltage measurement with improved cross-
talk attenuation, no compensation of measuring
lead resistance, with probe (3-pole) or without
probe (2-pole) as per IEC/EN 61557-5
Open circuit voltage: 23 to 24 V ac
Short circuit current: >50 mA ac
Measuring frequency: 128 Hz
Maximum permissible
overload:
250 Veff