1623-2
Users Manual
10
Description of Functions
The functions are selected with the central rotary switch. Measurement values are shown
on a liquid crystal display with correct decimal point and unit. Additional special
characters indicate measurement mode, operating condition, and error messages.
The Tester includes these measurement functions:
Earthing Resistance (R
E
) The earthing resistance is determined by a 3-pole or
4-pole current and voltage measurement. The
measuring voltage is a square pulse ac voltage with
48 / 20 V and a frequency of 94, 105, 111 or 128 Hz.
The frequency can be selected manually or
automatically (AFC).
Selective Measurement of
Earthing (R
E
)
Measurement of a single earth electrode in a mesh
operated (parallel) earthing system. The current flowing
through the single earth electrode is measured with an
external current transformer.
Low Battery Indicator Battery voltage is low, replace batteries.

Other manuals for Fluke 1623-2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 1623-2 and is the answer not in the manual?

Fluke 1623-2 Specifications

General IconGeneral
USB portYes
Control typeButtons, Rotary
CertificationIEC61326-1, CE
Housing colorBlack, Yellow
Display digits1999 digits
Measuring time6 s
Maximum voltage48 V
Display technologyLCD
Number of memory records1500
Storage temperature (T-T)-30 - 60 °C
Operating temperature (T-T)-10 - 50 °C
International Protection (IP) codeIP56
Battery sizeAA
Battery voltage1.5 V
Battery technologyAlkaline
Battery life (measurements)3000
Number of batteries supported6
Weight and Dimensions IconWeight and Dimensions
Depth187 mm
Width250 mm
Height133 mm
Weight1100 g