EasyManuals Logo

Fluke 17B MAX Calibration Manual

Fluke 17B MAX
14 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #13 background imageLoading...
Page #13 background image
Digital Multimeters
Adjustment Procedures
9
4. To put the DUT in adjustment mode, use a small probe to push the calibration button
(CAL1). See Figure 6.
Figure 6. Adjustment
5. Enter the adjustment values on the calibrator.
6. For each function, wait 4 seconds for the measurement to stabilize. Then push
H to
confirm and go to the next step. The display shows the function and the high voltage
indicator, see Ta b l e 4 .
7. For 17B MAX temperature adjustment:
a. Use an adjustable potentiometer to make a manual adjustment. See Figure 7.
b. Adjust to 0.1
ÁC to 0.3 ÁC at 0 ÁC input.
Tabl e 4. D ispl ay I ndic ato rs
A
High voltage alert
B
Function
CAL1
WP6
HOLD1
WP7
WP8
WP9
CAL1
1
2

Other manuals for Fluke 17B MAX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 17B MAX and is the answer not in the manual?

Fluke 17B MAX Specifications

General IconGeneral
BrandFluke
Model17B MAX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals