Digital Multimeter 
  Calibration Adjustments 
 17 
5.   in the display shows the CAL mode is set and the Meter is at the first calibration 
adjustment step. 
WCaution 
To prevent damage to the Meter, make sure the calibrator is in 
standby before you change its parameters, or change the 
function on the Meter. 
6.  Set the calibrator to the parameters shown in the Input Value column of Table 6 for 
the calibration step shown in the display. 
Note 
You can push  to show the necessary input signal level and G to 
show the necessary input signal frequency for the calibration step. 
8.  Turn on the output signal on the calibrator. 
9. Push D to complete the step and move to the subsequent step. 
If the step shown in the display moves to the subsequent step, the adjustment was 
successful. If the Meter gives two beeps and does not move to the subsequent step, 
then the adjustment was not successful. Make sure the calibrator is set correctly and 
push D again. If the calibrator output is correct, then it is necessary to repair the 
Meter. 
Note 
If the calibration adjustment is not completed correctly, the Meter will not 
operate correctly. 
10. Put the calibrator in standby. 
11. Do steps 6 through 10 for each calibration step in Table 6. 
12. After you complete the last adjustment step,  shows in the display. Push D to 
complete the calibration adjustment procedure and store the calibration constants in 
the Meter.