28 II Ex 
Calibration Information 
16 
Table 5. Button Functions for Calibration Adjustment 
Button CAL Function 
Password 
Numeric 
Value 
 
(Yellow) 
Push and hold to show the measured value in the display. The measurement 
is not calibrated, so it can be inaccurate. 
1 
 
Use to set the Meter to the CAL mode. In a calibration procedure, push and 
hold to show the necessary input level in the Meter display. 
2 
C  Push at the point where you type in the password to make a new password.  3 
D 
Use as the “ENTER” key. Push to store the new calibration adjustment value 
and move to the subsequent step. This button is also used to exit the 
calibration adjustment mode after the calibration adjustment sequence is 
complete. 
4 
H  Use only as a password value.  5 
E  Use only as a password value.  6 
F  Use only as a password value.  7 
G 
Push and hold to show the necessary frequency of the input signal in the 
Meter display. 
8 
Calibration Adjustment Procedure 
Note 
In the calibration adjustment procedure, some adjustment steps are longer 
to do than others (10 seconds to 15 seconds). For some of the steps, the 
Meter gives two beeps when the step is complete. Not all steps have this 
feature. 
1.  Push and hold down  while you turn the rotary switch from OFF to K. Z  
in the display shows you have set the Meter into the CAL mode. 
2. Push D once to see the number of calibrations that have been completed.  
3. Push D again to show  in the display.  
4.  Use the eight Meter buttons to type in the current password and push D. 
Note 
The default factory password is 1234. See the “How to Change the 
Password” section to change the password. If the password you typed in is 
not correct, then the Meter will beep two times and show   in the 
display.