Fluke 287 - AC Current Specifications

Fluke 287
38 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
287/289
Getting Started
8
AC Current Specifications
Accuracy
Function Range Resolution
20 to 45 Hz 45 to 1 kHz 1 to 20 kHz 20 to 100 kHz
[4]
500 μA 0.01 μA
1 % + 20 0.6 % + 20 0.6 % + 20 5 % + 40
AC μA
[3]
5000 μA 0.1 μA
1 % + 5 0.6 % + 5 0.6 % + 10 5 % + 40
50 mA 0.001 mA 1 % + 20 0.6 % + 20 0.6 % + 20 5 % + 40 AC mA
[3]
400 mA 0.01 mA 1 % + 5 0.6 % + 5 1.5 % + 10 5 % + 40
5 A 0.0001 A 1.5 % + 20 0.8 % + 20 3 % + 40
[4]
Not Spec’d AC A
[2]
10 A
[1]
0.001 A 1.5 % + 5 0.8 % + 5 3 % + 10
[4]
Not Spec’d
[1] 10 A range (10 % to 100 % of range).
[2] 20 A for 30 seconds on, 10 minutes off. >10 A not specified.
[3] 400 mA continuous; 550 mA for 2 minutes on, 1 minute off.
[4] Verified by design and type tests.
See Detailed Specifications introduction for additional information.