True-rms Digital Multimeters 
  General Specifications   
 7 
AC Voltage Specifications 
Accuracy 
Function Range Resolution 
20 to 45 Hz  45 to 65 Hz  65 Hz to 10 kHz  10 to 20 kHz  20 to 100 kHz 
50 mV
[1]
  0.001 mV  1.5 % + 60  0.3 % + 25  0.4 % + 25  0.7 % + 40  3.5 % + 40
[6]
 AC mV
[5] 
 
500 mV  0.01 mV  1.5 % + 60  0.3 % + 25  0.4 % + 25  0.7 % + 40  3.5 % + 40 
5 V
[1]
  0.0001 V  1.5 % + 60  0.3 % + 25  0.4 % + 25  1.5 % + 40  3.5 % + 40 
50 V
[1]
  0.001 V  1.5 % + 60  0.3 % + 25  0.4 % + 25  0.7 % + 40  3.5 % + 40 
500 V
[1]
  0.01 V  1.5 % + 60  0.3 % + 25  0.4 % + 25  Not Spec’d  Not Spec’d 
AC V 
1000 V  0.1 V  1.5 % + 60  0.3 % + 25  0.4 % + 25  Not Spec’d  Not Spec’d 
-70 to -62 dB
[3]
  0.01 dB  3 dB  1.5 dB  2 dB  2 dB  3 dB 
-62 to -52 dB
[3]
  0.01 dB  1.5 dB  1.0 dB  1 dB  1 dB  2 dB 
-52 to -6 dB
[3]
  0.01 dB  0.2 dB  0.1 dB  0.1 dB  0.2 dB  0.8 dB 
-6 to +34 dB
[3]
  0.01 dB  0.2 dB  0.1 dB  0.1 dB  0.2 dB  0.8 dB 
dBV 
34 to 60 dB
[3]
  0.01 dB  0.2 dB  0.1 dB  0.1 dB  Not Spec’d  Not Spec’d 
Low pass 
filter
[4]
 
 
2 % + 80  2 % + 40 
2 % +10
[2] 
-6 % -60
 
Not Spec’d  Not Spec’d 
L
[4]
 
1000 V  0.1 V  2 % + 80  2 % + 40  2 % + 40  Not Spec’d  Not Spec’d 
[1]  Below 5 % of range, add 20 counts. 
[2]  Specification increases linearly from -2 % at 200 Hz to -6 % at 440 Hz. Range is limited to 440 Hz. 
[3] dBm (600 Ω) is specified by adding +2.2 dB to the dBV range values. 
[4] 289 only. 
[5]  500 mV range specification between 64.000 kHz and 67.000 kHz is +(0.0 % to -5 % of reading); 50 mV range specification between 64.000 kHz and 
67.000 kHz is +(0.0 % to -6 % of reading ±40 counts). Valid from -20 °C to +55 °C. 
[6]  Add 2 % above 75 kHz. 