Getting Started
16
MIN MAX, Recording, and Peak Specifications
Function Nominal Response Accuracy
200 ms to 80% (dc function)
Specified accuracy ±12 counts for changes >425 ms in duration in manual
range.
MIN MAX, Recording
350 ms to 80 % (ac function)
Specified Accuracy ±40 counts for changes >1.5 s in duration in manual
range.
Peak 250 μS (peak)
[1]
Specified accuracy ±100 counts up to 5,000 count (full range) reading. For
higher peak reading (to 12,000 counts), specified accuracy ±2 % of
reading.
[1] For repetitive peaks; 2.5 ms for single events.
Burden Voltage (A, mA,
μ
A)
Function Range Burden Voltage
500 μA 102 μV/ μA
5000 μA 102 μV/ μA
50.000 mA 1.8 mV/mA
mA, μA
400.00 mA 1.8 mV/mA
5.0000 A 0.04 V/A A
10.000 A 0.04 V/A

Other manuals for Fluke 289

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 289 and is the answer not in the manual?

Fluke 289 Specifications

General IconGeneral
Product typeAnalog multimeter
AC current range5 - 10 A
Continuity checkYes
DC voltage range5 - 1000 V
Resistance range50 - 500 Ω
Capacitance range1 - 100 nF
Resolution (frequency)999.99 Hz
Resolution (AC current)5000 µA
Resolution (DC voltage)500 mV
Temperature measurement range-20 - 55 °C
Dimensions (WxDxH)10.2 x 6 x 22.2 mm
Battery voltage1000 V
Weight and Dimensions IconWeight and Dimensions
Weight870.9 g

Related product manuals