287/289 
Getting Started 
10 
DC Current Specifications 
Accuracy 
AC over DC, DC over AC, AC + DC
[1]
 
Function Range Resolution 
DC
[1][3]
 
20 to 45 Hz  45 Hz to 1 kHz  1 to 20 kHz  20 to 100 kHz
[5]
500 μA 0.01 μA 
0.075 % + 20  1 % + 20  0.6 % + 20  0.6 % + 20  5 % + 40 
DC μA
[4]
 
5000 μA 0.1 μA 
0.075 % + 2  1 % + 5  0.6 % + 5  0.6 % + 10  5 % + 40 
50 mA  0.001 mA  0.05 % + 10
[6]
  1 % + 20  0.6 % + 20  0.6 % + 20  5 % + 40 DC mA
[4]
 
400 mA  0.01 mA  0.15 % + 2  1 % + 5  0.6 % + 5  1.5 % + 10  5 % + 40 
5 A  0.0001 A  0.3 % + 10  1.5 % + 20  0.8 % + 20  3 % + 40
[5]
 Not Spec’d DC A
[2]
 
10 A  0.001 A  0.3 % + 2  1.5 % + 10  0.8 % + 10  3 % + 10
[5]
 Not Spec’d 
[1]  AC+DC ranges are specified from 2 % to 140 % of range. 
[2]  20 A for 30 seconds on, 10 minutes off. >10 A not specified. 
[3]  Add 20 counts in dual display ac over dc, dc over ac or ac+dc. 
[4]  400 mA continuous; 550 mA for 2 minutes on, 1 minute off. 
[5]  Verified by design and type tests. 
[6] Temperature coefficient: 0.1 X (specified accuracy)/ °C (<18 °C or > 28 °C)