Power Quality Clamp Meter
Calibration Verification Procedure
25
Amps Check
Mode: Amps
Input: dc & 60/50 Hz ac signal
Internal Accuracy: ± 1.5 % rdg ± 5 dgts for A > 10.0 A
± 0.2 A for A 10.0 A
Harmonic Distortion
Harmonic distortion parameters are not checked because they use the same hardware
paths as are used in measurement of amps and volts.
Frequency Check
Mode: Volts Frequency
Input: 2 V 60/50 Hz
Internal Accuracy: ± 0.5 % rdg 40 - 70 Hz
Calibrator Specification
(Amps) Min Max
4.00 3.80 4.20
10.00 9.80 10.20
11.00 10.79 11.22
39.00 38.37 39.64
41.0 39.9 42.1
390.0 383.7 396.4
410 399 421
1000 980 1020