Multi-Product Calibrator
Additional Specifications
23
AC Voltage (Non-Sine Wave) (cont.)
Square Wave
Range (p-p)
[1]
Frequency
1-Year Absolute Uncertainty, tcal
±5 °C, ±(% of output + % of range)
[2]
Max Voltage Resolution
Normal Channel (Single Output Mode)
2.9 to 65.999 mV
0.01 to 10 Hz 5.0 + 0.5 Two digits on each range
10 to 45 Hz 0.25 + 0.5
Six digits on each range
45 Hz to 1 kHz 0.25 + 0.25
1 to 20 kHz 0.5 + 0.25
20 to 100 kHz 5.0 + 0.5
66 to 659.999 mV
0.01 to 10 Hz 5.0 + 0.5 Two digits on each range
10 to 45 Hz 0.25 + 0.5
Six digits on each range
45 Hz to 1 kHz 0.25 + 0.25
1 to 20 kHz 0.5 + 0.25
20 to 100 kHz 5.0 + 0.5
0.66 to 6.59999 V
0.01 to 10 Hz 5.0 + 0.5 Two digits on each range
10 to 45 Hz 0.25 + 0.5
Six digits on each range
45 Hz to 1 kHz 0.25 + 0.25
1 to 20 kHz 0.5 + 0.25
20 to 100 kHz 5.0 + 0.5
6.6 to 66.0000 V
0.01 to 10 Hz 5.0 + 0.5 Two digits on each range
10 to 45 Hz 0.25 + 0.5
Six digits on each range
45 Hz to 1 kHz 0.25 + 0.25
1 to 20 kHz 0.5 + 0.25
20 to 100 kHz 5.0 + 0.5
45 Hz to 1 kHz 0.25 + 0.25
1 to 10 kHz
[3]
5.0 + 0.5
[1] To convert p-p to rms for square wave, multiply the p-p value by 0.5.
[2] Uncertainty is stated in p-p. Amplitude is verified using an rms-responding DMM.
AC Voltage, DC Offset
Range
[1]
(Normal Channel) Offset Range
[2]
Max Peak
Signal
1-Year Absolute Uncertainty, tcal
±5 °C
[3]
±(% of dc output + floor)
Sine Waves (rms)
3.3 to 32.999 mV 0 to 50 mV 80 mV
0.1 + 33 μV
33 to 329.999 mV 0 to 500 mV 800 mV
0.1 + 330 μV
0.33 to 3.29999 V 0 to 5 V 8 V
0.1 + 3300 μV
3.3 to 32.9999 V 0 to 50 V 55 V 0.1 + 33 mV
Triangle Waves and Truncated Sine Waves (p-p)
9.3 to 92.999 mV 0 to 50 mV 80 mV
0.1 + 93 μV
93 to 929.999 mV 0 to 500 mV 800 mV
0.1 + 930 μV
0.93 to 9.29999 V 0 to 5 V 8 V
0.1 + 9300 μV
9.3 to 93.0000 V 0 to 50 V 55 V 0.1 + 93 mV
Square Waves (p-p)
6.6 to 65.999 mV 0 to 50 mV 80 mV
0.1 + 66 μV
66 to 659.999 mV 0 to 500 mV 800 mV
0.1 + 660 μV
0.66 to 6.59999 V 0 to 5 V 8 V
0.1 + 6600 μV
6.6 to 66.0000 V 0 to 50 V 55 V 0.1 + 66 mV
[1] Offsets are not allowed on ranges above the highest range shown above.
[2] The maximum offset value is determined by the difference between the peak value of the selected voltage output and the
allowable maximum peak signal. For example, a 10 V p-p square wave output has a peak value of 5 V, allowing a maximum
offset up to ± 50 V to not exceed the 55 V maximum peak signal. The maximum offset values shown above are for the
minimum outputs in each range.
[3] For frequencies 0.01 to 10 Hz, and 500 kHz to 2 MHz, the offset uncertainty is 5 % of output, ±1 % of the offset range.