Multi-Product Calibrator
Performance Tests
37
Table 12. Verification Tests for Frequency
Range, Normal
Output, V
Output, Normal, V Frequency Lower Limit [1] Upper Limit [1]
3.29999 3.00000
119.00 Hz 118.99602 Hz 119.00398 Hz
120.0 Hz 119.99600 Hz 120.00400 Hz
1000.0 Hz 999.974000 Hz 1000.026000 Hz
100.00 kHz 99.99750000 Hz 100.00250000 Hz
[1] Frequency accuracy is specified for 1 year.