35
Table 9. Verification Tests for Capacitance
Range Output
Test Frequency
or Current
Lower Limit Upper Limit
0.3999 nF 0.2200 nF 5 kHz 0.2192 nF 0.2308 nF
0.3999 nF 0.3500 nF 1 kHz 0.3387 nF 0.3613 nF
1.0999 nF 0.4800 nF 1 kHz 0.4682 nF 0.4918 nF
1.0999 nF 0.6000 nF 1 kHz 0.5877 nF 0.6123 nF
1.0999 nF 1.0000 nF 1 kHz 0.9862 nF 1.0138 nF
3.299 nF 2.0000 nF 1 kHz 1.9824 nF 2.0176 nF
10.999 nF 7.0000 nF 1 kHz 6.9767 nF 7.0233 nF
10.999 nF 10.9000 nF 1 kHz 10.8693 nF 10.9307 nF
32.999 nF 20.000 nF 1 kHz 19.8620 nF 20.1380 nF
109.99 nF 70.00 nF 1 kHz 69.767 nF 70.233 nF
109.99 nF 109.00 nF 1 kHz 108.693 nF 109.307 nF
329.99 nF 200.00 nF 1 kHz 199.320 nF 200.680 nF
329.99 nF 300.00 nF 1 kHz 299.130 nF 300.870 nF
1.0999 μF 0.7000 μF 100 Hz 0.69767 μF 0.70233 μF
1.0999 μF 1.0900 μF 100 Hz 1.05929 μF. 1.12071 μF
3.2999 μF 2.0000 μF 100 Hz 1.99320 μF 2.00680 μF
3.2999 μF 3.0000 μF 100 Hz 2.99130 μF 3.00870 μF
10.999 μF 7.000 μF 100 Hz 6.9767 μF 7.0233 μF
10.999 μF 10.900 μF 100 Hz 10.8693 μF 10.9307 μF
32.999 μF 20.000 μF 100 Hz 19.9100 μF 20.0900 μF
32.999 μF 30.000 μF 100 Hz 29.8800 μF 30.1200 μF
109.99 μF 70.00 μF 50 Hz 69.662 μF 70.338 μF
109.99 μF 109.00 μF 50 Hz 108.529 μF 109.471 μF
329.99 μF 200.00 μF 54 μA dc 199.020 μF 200.980 μF
329.99 μF 300.00 μF 80 μA dc 298.680 μF 301.320 μF
1.0999 mF 0.3300 mF 90 μA dc 0.32788 mF 0.33212 mF
1.0999 mF 0.7000 mF 180 μA dc 0.69662 mF 0.70338 mF
1.0999 mF 1.0900 mF 270 μA dc 1.08529 mF 1.09471 mF
3.299 mF 1.100 mF 270 μA dc 1.0933 mF 1.1067 mF