5502E
Getting Started
24
AC Voltage, Square Wave Characteristics
Risetime @
1 kHz
Typical
Settling Time @
1 kHz Typical
Overshoot
@ 1 kHz
Typical
Duty Cycle Range Duty Cycle Uncertainty
<1 μs
<10 μs to 1 % of
final value
<2 %
1 % to 99 % <3.3 V p-p.
0,01 Hz to 100 kHz
±(0.02 % of period + 100 ns), 50 % duty cycle
±(0.05 % of period + 100 ns), other duty cycles
from 10 % to 90 %
AC Voltage, Triangle Wave Characteristics (typical)
Linearity to 1 kHz Aberrations
0.3 % of p-p value, from 10 % to 90 % point <1 % of p-p value, with amplitude >50 % of range
AC Current (Non-Sine Wave)
Triangle Wave &
Truncated Sine Wave
Range p-p
Frequency
1-Year Absolute Uncertainty tcal ±5 °C
±(% of output + % of range)
Max Current
Resolution
0.047 to 0.92999 mA
[1]
10 to 45 Hz 0.25 + 0.5
Six digits
45 Hz to 1 kHz 0.25 + 0.25
1 to 10 kHz 10 + 2
0.93 to 9.29999 mA
[1]
10 to 45 Hz 0.25 + 0.5
Six digits
45 Hz to 1 kHz 0.25 + 0.25
1 to 10 kHz 10 + 2
9.3 to 92.9999 mA
[1]
10 to 45 Hz 0.25 + 0.5
Six digits
45 Hz to 1 kHz 0.25 + 0.25
1 to 10 kHz 10 + 2
93 to 929.999 mA
[1]
10 to 45 Hz 0.25 + 0.5
Six digits
45 Hz to 1 kHz 0.25 + 0.5
1 to 10 kHz 10 + 2
0.93 to 8.49999 A
[2]
10 to 45 Hz 0.5 + 1.0
Six digits
45 Hz to 1 kHz 0.5 + 0.5
1 to 10 kHz 10 + 2
8.5 to 57 A
[2]
45 to 500 Hz 0.5 + 0.5
500 Hz to 1 kHz 1.0 + 1.0