EasyManua.ls Logo

Fluke 5502E - AC Voltage, Square Wave Characteristics; AC Voltage, Triangle Wave Characteristics (Typical); AC Current (Non-Sine Wave)

Fluke 5502E
48 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5502E
Getting Started
24
AC Voltage, Square Wave Characteristics
Risetime @
1 kHz
Typical
Settling Time @
1 kHz Typical
Overshoot
@ 1 kHz
Typical
Duty Cycle Range Duty Cycle Uncertainty
<1 μs
<10 μs to 1 % of
final value
<2 %
1 % to 99 % <3.3 V p-p.
0,01 Hz to 100 kHz
±(0.02 % of period + 100 ns), 50 % duty cycle
±(0.05 % of period + 100 ns), other duty cycles
from 10 % to 90 %
AC Voltage, Triangle Wave Characteristics (typical)
Linearity to 1 kHz Aberrations
0.3 % of p-p value, from 10 % to 90 % point <1 % of p-p value, with amplitude >50 % of range
AC Current (Non-Sine Wave)
Triangle Wave &
Truncated Sine Wave
Range p-p
Frequency
1-Year Absolute Uncertainty tcal ±5 °C
±(% of output + % of range)
Max Current
Resolution
0.047 to 0.92999 mA
[1]
10 to 45 Hz 0.25 + 0.5
Six digits
45 Hz to 1 kHz 0.25 + 0.25
1 to 10 kHz 10 + 2
0.93 to 9.29999 mA
[1]
10 to 45 Hz 0.25 + 0.5
Six digits
45 Hz to 1 kHz 0.25 + 0.25
1 to 10 kHz 10 + 2
9.3 to 92.9999 mA
[1]
10 to 45 Hz 0.25 + 0.5
Six digits
45 Hz to 1 kHz 0.25 + 0.25
1 to 10 kHz 10 + 2
93 to 929.999 mA
[1]
10 to 45 Hz 0.25 + 0.5
Six digits
45 Hz to 1 kHz 0.25 + 0.5
1 to 10 kHz 10 + 2
0.93 to 8.49999 A
[2]
10 to 45 Hz 0.5 + 1.0
Six digits
45 Hz to 1 kHz 0.5 + 0.5
1 to 10 kHz 10 + 2
8.5 to 57 A
[2]
45 to 500 Hz 0.5 + 0.5
500 Hz to 1 kHz 1.0 + 1.0
[1] Frequency limited to 1 kHz with LCOMP on.
[2] Frequency limited to 440 Hz with LCOMP on.

Related product manuals