EasyManua.ls Logo

Fluke 5502E - AC Voltage Verification; AC Voltage (Normal) Tests

Fluke 5502E
48 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5502E
Getting Started
30
Table 7. Verification Tests for AC Voltage (Normal)
Range Output Frequency Lower Limit Upper Limit
32.999 mV 3.000 mV 45 Hz 2.977 mV 3.022 mV
32.999 mV 3.000 mV 10 kHz 2.977 mV 3.022 mV
32.999 mV 30.000 mV 9.5 Hz 28.350 mV 31.650 mV,
32.999 mV 30.000 mV 10 Hz 29.944 mV 30.056 mV
32.999 mV 30.000 mV 45 Hz 29.956 mV 30.044 mV
32.999 mV 30.000 mV 1 kHz 29.956 mV 30.044 mV
32.999 mV 30.000 mV 10 kHz 29.956 mV 30.044 mV
32.999 mV 30.000 mV 20 kHz 29.944 mV 30.056 mV
32.999 mV 30.000 mV 50 kHz 29.932 mV 30.068 mV
32.999 mV 30.000 mV 100 kHz 29.877 mV 30.123 mV
32.999 mV 30.000 mV 450 kHz 29.715 mV 30.285 mV
329.999 mV 33.000 mV 45 Hz 32,970 mV 33.029 mV
329.999 mV 33.000 mV 10 kHz 32.970 mV 33.029 mV
329.999 mV 300.000 mV 9.5 Hz 283.350 mV 316.650 mV
329.999 mV 300.000 mV 10 Hz 299.917 mV 300.083 mV
329.999 mV 300.000 mV 45 Hz 299.893 mV 300.107 mV
329.999 mV 300.000 mV 1 kHz 299.983 mV 300.107 mV
329.999 mV 300.000 mV 10 kHz 299.983 mV 300.107 mV
329.999 mV 300.000 mV 20 kHz 299.782 mV 300.218 mV
329.999 mV 300.000 mV 50 kHz 299.702 mV 300.298 mV
329.999 mV 300.000 mV 100 kHz 299.311 mV 300.689 mV
329.999 mV 300.000 mV 500 kHz 298.470 mV 301.530 mV
3.29999 V 0.33000 V 45 Hz 0.32984 V 0.33015 V
3.29999 V 0.33000 V 10 kHz 0.32984 V 0.33015V
3.29999 V 3.00000 V 9.5 Hz 2.83500 V 3.16500 V
3.29999 V 3.00000 V 10 Hz 2.99868 V 3.00132 V
3.29999 V 3.00000 V 45 Hz 2.99910 V 3.00090 V
3.29999 V 3.00000 V 1 kHz 2.99910V 3.00090 V
3.29999 V 3.00000 V 10 kHz 2.99910 V 3.00090 V
3.29999 V 3.00000 V 20 kHz 2.99817 V 3.00183 V
3.29999 V 3.00000 V 50 kHz 2.99745 V 3.00255 V