Table 7. Verification Tests for AC Voltage (Normal) (cont.)
Range Output Frequency Lower Limit Upper Limit
3.29999 V 3.00000 V 100 kHz 2.99437 V 3.00563V
3.29999 V 3.00000 V 450 kHz 2.98659 V 3.01340 V
3.29999 V 3.29000 V 1 MHz 2.250 V
[1]
32.9999 V 3.3000 V 45 Hz 3.2985 V 3.3014 V
32.9999 V 3.3000 V 10 kHz 3.2985 V 3.3014 V
32.9999 V 30.0000 V 9.5 Hz 28.3500 V 31.6500 V
32.9999 V 30.0000 V 10 Hz 29.9866 V 30.0134V
32.9999 V 30.0000 V 45 Hz 29.9919 V 30.0081 V
32.9999 V 30.0000 V 1 kHz 29.9919 V 30.0081 V
32.9999 V 30.0000 V 10 kHz 29.9919 V 30.0081 V
32.9999 V 30.0000 V 20 kHz 29.9802 V 30.0198 V
32.9999 V 30.0000 V 50 kHz 29.9736 V 30.0264 V
32.9999 V 30.0000 V 90 kHz 29.9404 V 30.0596 V
329.999 V 33.000 V 45 Hz 32.984 V 33.015 V
329.999 V 33.000 V 10 kHz 32.969 V 33.030V
329.999 V 300.000 V 45 Hz 299.880 V 300.120 V
329.999 V 300.000 V 1 kHz 299.880 V 300.120 V
329.999 V 300.000 V 10 kHz 299.799 V 300.201 V
329.999 V 300.000 V 18 kHz 299.754 V 300.246 V
329.999 V 300.000 V 50 kHz 299.703 V 300.297 V
329.999 V 200.000 V 100 kHz 199.536 V 200.464 V
1020.00 V 330.00 V 45 Hz 329.84 V 330.15 V
1020.00 V 330.00 V 10 kHz 329.73 V 330.26 V
1020.00 V 1000.00V 45 Hz 999.56 V 1000.44 V
1020.00 V 1000.00 V 1 kHz 999.56 V 1000.44 V
1020.00 V 1000.00 V 5 kHz 999.349 V 1000.66 V
1020.00 V 1000.00 V 8 kHz 999.23 V 1000.77 V
1020.00 V 1020.00 V 1 kHz 1019.55 V 1020.44 V
1020.00 V 1020.00 V 8 kHz 1019.21 V 1020.78 V
[1] Typical specification is -24 dB at 2 MHz