Multi-Product Calibrator
Performance Tests
33
Table 8. Verification Tests for AC Current (cont.)
Range Output Frequency Lower Limit Upper Limit
32.999 mA 19.0000 mA 30 kHz 18.935 mA 19.065 mA
32.999 mA 32.9000 mA 10 Hz 32.849 mA 32.951 mA
32.999 mA 32.9000 mA 1 kHz 32.886 mA 32.914 mA
32.999 mA 32.9000 mA 5 kHz 32.877 mA 32.923 mA
32.999 mA 32.9000 mA 10 kHz 32.844 mA 32.956 mA
32.999 mA 32.9000 mA 30 kHz 32.791 mA 33.009 mA
329.99 mA 33.0000 mA 1 kHz 32.97 mA 33.03 mA
329.99 mA 33.0000 mA 5 kHz 32.92 mA 33.08 mA
329.99 mA 33.0000 mA 30 kHz 32.69 mA 33.31 mA
329.99 mA 190.0000 mA 1 kHz 189.91 mA 190.09 mA
329.99 mA 190.0000 mA 10 kHz 189.60 mA 190.40 mA
329.99 mA 190.0000 mA 30 kHz 189.19 mA 190.81 mA
329.99 mA 329.0000 mA 10 Hz 328.49 mA 329.51 mA
329.99 mA 329.0000 mA 45 Hz 328.86 mA 329.14 mA
329.99 mA 329.0000 mA 1 kHz 328.86 mA 329.14 mA
329.99 mA 329.0000 mA 5 kHz 328.69 mA 329.31 mA
329.99 mA 329.0000 mA 10 kHz 328.37 mA 329.63 mA
329.99 mA 329.0000 mA 30 kHz 327.75 mA 330.25 mA
2.99999 A 0.33000 A 1 kHz 0.32978 A 0.33022 A
2.99999 A 0.33000 A 5 kHz 0.32735 A 0.33265 A
2.99999 A 0.33000 A 10 kHz 0.31840 A 0.34160 A
2.99999 A 1.09000 A 10 Hz 1.08827 A 1.09174 A
2.99999 A 1.09000 A 45 Hz 1.08951 A 1.09049 A
2.99999 A 1.09000 A 1 kHz 1.08951 A 1.09049 A
2.99999 A 1.09000 A 5 kHz 1.08355 A 1.09645 A
2.99999 A 1.09000 A 10 kHz 1.06320 A 1.11680A
2.99999 A 2.99000 A 10 Hz 2.98542 A 2.99459 A
2.99999 A 2.99000 A 45 Hz 2.98840 A 2.99160 A
2.99999 A 2.99000 A 1 kHz 2.98840 A 2.99160 A
2.99999 A 2.99000 A 5 kHz 2.97405 A 3.00595 A