5700A/5720A Series II Calibrator
Service Manual
1-28
1-31. Resistance Specifications
5720A Series II Resistance Specifications
Absolute Uncertainty of Characterized Value
± 5 °C from calibration temperature
[1]
Relative Uncertainty
± 1 °C
24 Hours 90 Days 180 Days 1 Year 24 Hours 90 Days
Nominal Value
(Ω)
±ppm
99 % Confidence Level
0
1
1.9
10
19
100
190
1 k
1.9 k
10 k
19 k
100 k
190 k
1 M
1.9 M
10 M
19 M
100 M
50 μΩ
85
85
23
23
10
10
8
8
8
9
9
9
16
17
33
43
100
50 μΩ
95
95
25
25
11
11
9
9
9
9
11
11
18
19
37
47
110
50 μΩ
100
100
26
26
11.5
11.5
9.5
9.5
9.5
9.5
12
12
20
21
40
50
115
50 μΩ
110
110
27
27
12
12
10
10
10
10
13
13
23
24
46
55
120
50 μΩ
32
25
5
4
2
2
2
2
2
2
2
2
2.5
3
10
20
50
50 μΩ
40
33
8
7
4
4
3
3
3
3
3
3
5
6
14
24
60
95 % Confidence Level
0
1
1.9
10
19
100
190
1 k
1.9 k
10 k
19 k
100 k
190 k
1 M
1.9 M
10 M
19 M
100 M
40 μΩ
70
70
20
20
8
8
6.5
6.5
6.5
7.5
7.5
7.5
13
14
27
35
85
40 μΩ
80
80
21
21
9
9
7.5
7.5
7.5
7.5
9
9
15
16
31
39
95
40 μΩ
85
85
22
22
9.5
9.5
8
8
8
8
10
10
17
18
34
42
100
40 μΩ
95
95
23
23
10
10
8.5
8.5
8.5
8.5
11
11
20
21
40
47
100
40 μΩ
27
20
4
3.5
1.6
1.6
1.6
1.6
1.6
1.6
1.6
1.6
2
2.5
8
16
40
40 μΩ
35
26
7
6
3.5
3.5
2.5
2.5
2.5
2.5
2.5
2.5
4
4
12
20
50
Note:
1. Specifications apply to displayed value. 4-wire connections, except 100 MΩ.