EasyManua.ls Logo

Fluke 5700A Series II - Resistance Specifications

Fluke 5700A Series II
186 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Introduction and Specifications
Electrical Specifications 1
1-29
5700A Series II Resistance Specifications
Absolute Uncertainty of Characterized Value
± 5 °C from calibration temperature
[1]
Relative Uncertainty
± 1 °C
24 Hours 90 Days 180 Days 1 Year 24 Hours 90 Days
Nominal Value
(Ω)
±ppm
99 % Confidence Level
0
1
1.9
10
19
100
190
1 k
1.9 k
10 k
19 k
100 k
190 k
1 M
1.9 M
10 M
19 M
100 M
50 μΩ
85
85
26
24
15
15
11
11
9
9
11
11
16
17
33
43
110
50 μΩ
95
95
28
26
17
17
12
12
11
11
13
13
18
19
37
47
120
50 μΩ
100
100
30
28
18
18
13
13
12
12
14
14
20
21
40
50
125
50 μΩ
110
110
33
31
20
20
15
15
14
14
16
16
23
24
46
55
130
50 μΩ
32
25
5
4
2
2
2
2
2
2
2
2
2.5
3.5
10
20
50
50 μΩ
40
33
8
7
4
4
3.5
3.5
3.5
3.5
3.5
3.5
5
6
14
24
60
95 % Confidence Level
0
1
1.9
10
19
100
190
1 k
1.9 k
10 k
19 k
100 k
190 k
1 M
1.9 M
10 M
19 M
100 M
50 μΩ
70
70
21
20
13
13
9
9
7.5
7.5
9
9
13
14
27
35
90
50 μΩ
80
80
23
22
14
14
10
10
9.5
9.5
11
11
15
16
31
39
100
50 μΩ
85
85
27
24
15
15
11
11
10.5
10.5
12
12
17
18
34
42
105
50 μΩ
95
95
28
27
17
17
13
13
12
12
14
14
20
21
40
47
110
50 μΩ
32
25
5
4
2
2
2
2
2
2
2
2
2.5
3
10
20
50
50 μΩ
40
33
8
7
4
4
3.5
3.5
3.5
3.5
3.5
3.5
5
6
14
24
60
Note:
1. Specifications apply to displayed value. 4-wire connections, except 100 MΩ.